2. Warnings and Precautions
9
I
PaX-i3D Smart
English
❚
Emergency stop
● If a problem occurs during image acquisition, press the red emergency stop button
to immediately stop all moving parts and cut off all power to the equipment.
❚
Trouble-free operation
● Never use this equipment in an environment that is susceptible to explosion.
● Always operate the equipment within a temperature range of 10 °C to 35 °C (50 °F
to 95 °F) for the safe operation. Image quality may deteriorate if the equipment is
operated outside of this range.
● Always allow the equipment sufficient time to warm up (while switched on) if it has
been exposed to temperatures of below 10 °C (50 °F).
● Only perform X-rays of patients if system is in full working order.
● Always ensure that equipment movement is not obstructed by the patient’s clothing,
medical device (such as a wheelchair), or the patient themselves.
● Do not leave the patient unattended around the equipment.
● Remove all radio-controlled devices, mobile phones, etc. from the X-ray room prior
to image acquisition as these objects may cause the equipment to malfunction.
❚
Modifying the equipment
● Modifying the equipment in any way which may affect the safety of the operator,
patient or other persons is prohibited by law.
● No part of this equipment is serviceable by the user. All maintenance and repair of
this equipment must be performed by a qualified service technician.
● This product may only be operated with original VATECH accessories or third-party
accessories expressly approved by VATECH.
2.2
Electricity-related Safety Precautions
● Check the status of the power source, PC and cables prior to operating the
equipment.
● Ensure that the on/off switch is set to off when the equipment is not in use.
● Always disconnect the power supply before cleaning the equipment.
● Always keep electrical cords away from hot appliances or radiators.