13. Appendices
139
I
PaX-i3D Smart
English
13.3
Electromagnetic Compatibility (EMC) Information
Guidance and manufacturer’s declaration - electromagnetic emissions.
The model PaX-i3D Smart(PHT-30LFO) is intended for use in the electromagnetic environment
specified below. The customer or the user of the model PaX-i3D Smart(PHT-30LFO) should
assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The model PaX-i3D Smart(PHT-30LFO)
uses RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A (The model PaX-
i3D Smart(PHT-30LFO)
in combination with the
shielded location)
The model PaX-i3D Smart(PHT-30LFO)
must be used only in a shielded location
with a minimum RF shielding effectiveness
and, for each cable that exits the shielded
location, a minimum RF filter attenuation
of 20 dB from 30 MHz to 230 MHz, 20
dB from 230 MHz to 1 GHz. The model
PaX-i3D Smart(PHT-30LFO) , when
installed in such a shielded location, is
suitable for use in all establishments
other than domestic, and may be used in
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Not applicable
Voltage
fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
NOTE) It is essential that the actual RF shielding effectiveness and filter attenuation of the
shielded location be verified to ensure that they meet or exceed the specified minimum values.