13-2
IM CW240E
13.1 Analog Output (DA Output)
● Output (4CH)
The CW240 can convert measurement values to direct current voltage signals
and output them.
13.1.1 Analog Output Specifications
Output Voltage : ⫾1VFS of the rated value of each range
Output Current : ⫾1 mA maximum
Output Quantity: 4CH
Output Data : Selected from measurement items
Accuracy : ⫾(measurement accuracy + 0.5% F.S.)
Response : 3 times or less the input signal frequency
(excluding harmonic measurement)
16 times or less the input signal frequency
(harmonic measurement)
Update Cycle : 1 cycle of the input signal frequency
(excluding harmonic measurement)
16 cycles of the input signal frequency
(harmonic measurement)
● Response
<Other than harmonic measurement>
DA output is output after the input signal waveform.
The delay (response) is 3 times or less a single cycle of the input signal
frequency (3 cycles or less).
<Harmonic measurement>
DA output is output after the input signal waveform.
The delay (response) is 16 cycles or less of the input signal frequency.
Chapter 13 Using Analog Input/Output (Optional)