4-7
IM CW240E
Wiring
4
4.5 Carrying out Wiring
WARNING
• When attaching voltage probes to or clamping a clamp-on probe on the circuit
under test, turn off power to the circuit under test.
It is very dangerous to connect or disconnect a voltage probe or clamp or
unclamp a clamp-on probe without turning off the circuit under test.
• Be sure to connect voltage probes to or clamp a clamp-on probe on the sec-
ondary side of the circuit under test, such as current limiters (circuit break-
ers). Should an accident such as a short occur, other circuits will be protected
by these circuit breakers.
v
1
N
v
2
v
3
L
1
CH
H
L
H
L
H
V INPU
T
V INPUT
600V MAX
600V MAX
TERMINALS 600V MAX
TERMINALS 600V MAX
TO EAR
O EAR
TH
TH
L
2
H
L
CH
CH
1
H
L
CH
CH
CA
CA
T.
LOGIC NIPUT
LOGIC NIPUT
ST
ST
AR
AR
T/ST
T/ST
OP
OP
5.5V MAX
5.5V MAX
L
H
L
H
LOGIC NIPUT
LOGIC NIPUT
ST
ST
AR
AR
T/ST
T/ST
OP
OP
5.5V MAX
5.5V MAX
Power supply
side
Circuit breaker
Power supply side
Load side
Load side
Direction marks
Clip voltage probes securely onto
metal parts such as screws or
wiring bars.
Orientate the clamp-on
probe so that the
arrows on the clamping
CT faces the correct
direction.