3-5
IM DL350-17EN
Programming Overview
1
2
3
4
5
App
Index
3.3 Response
Response
When the controller sends a query with a question
mark, the instrument returns a response message to
the query. The instrument returns response messages
in one of the following two forms.
Response consisting of a header and data
Responses that can be used as program messages
without any changes are returned with command
headers attached.
Example
:ACQUire:MODE?<PMT>
->:ACQUire:MODE NORMAL<RMT>
Response only consisting of data
Responses that cannot be used as program
messages unless changes are made (query-only
commands) are returned without headers. However,
there are query-only commands whose responses
the instrument will attach headers to.
Example
:MEASure:CHANnel1:PTOPeak:
VALue?<PMT>
-> 10.0E+00<RMT>
If You Want the Instrument to Return
Responses without Headers
You can configure the instrument so that
even responses that have both headers and
data are returned without headers. Use the
COMMunicate:HEADer
command for this purpose.
Abbreviated Form
The instrument normally returns response headers with
the lower-case section removed. You can configure the
instrument so that full headers are returned. Use the
COMMunicate:VERBose
command for this purpose.
The sections enclosed in square brackets ([ ]) are also
omitted in the abbreviated form.

Other manuals for YOKOGAWA DL350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL350 and is the answer not in the manual?

YOKOGAWA DL350 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL350
CategoryRecording Equipment
LanguageEnglish

Related product manuals