EasyManuals Logo
Home>YOKOGAWA>Recording Equipment>DL350

YOKOGAWA DL350 User Manual

YOKOGAWA DL350
215 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
4-39
IM DL350-17EN
Commands
1
2
3
4
5
App
Index
4.5 CALibrate Group
The commands in this group deal with calibration. You can perform the same operations and make the same settings
and queries that you can by using Preparation > Calibration in the top menu.
:CALibrate?
Function Queries all calibration settings.
Syntax
:CALibrate?
:CALibrate[:EXECute]
Function Executes calibration.
This command is different from the common
command *CAL? in that this command does
not return the results when the calibration is
completed.
Syntax
:CALibrate[:EXECute]
:CALibrate:MODE
Function Sets or queries the auto calibration mode.
Syntax
CALibrate:MODE {AUTO|OFF}
CALibrate:MODE?
Example
:CALIBRATE:MODE AUTO
:CALIBRATE:MODE?
-> :CALIBRATE:MODE AUTO
:CALibrate:SBOCancel:CHANnel<x>[:
STATus] (Strain Balance & Offset
Cancel)
Function Sets or queries whether the specified channel
is included in the balancing of all channels on
strain modules or the DC offset canceling of all
channels on voltage measurement modules.
Syntax
:CALibrate:SBOCancel:CHANnel<x>[:
STATus] {<Boolean>}
:CALibrate:SBOCancel:CHANnel<x>[:
STATus]?
<x> = 1 to 4
Example
:CALIBRATE:SBOCANCEL:CHANNEL1 ON
:CALIBRATE:SBOCANCEL:CHANNEL1?
-> :CALIBRATE:SBOCANCEL:CHANNEL1 1
Description This command is valid when any of the following
modules is installed.
701270(STRAIN_NDIS),
701271(STRAIN_DSUB), 720211 (HS100M12),
720250 (HS10M12)
:CALibrate:SBOCancel:CHANnel<x1>:SCH
annel<x2>[:STATus] (Strain Balance &
Offset Cancel)
Function Sets or queries whether the specified 4-CH
module’s channel is included in the balancing of
all channels on strain modules or the DC offset
canceling of all channels on voltage measurement
modules.
Syntax
:CALibrate:SBOCancel:CHANnel<x1>:
SCHannel<x2>[:STATus] {<Boolean>}
:CALibrate:SBOCancel:CHANnel<x1>:
SCHannel<x2>[:STATus]?
<x1> = 1 to 4
<x2> = 1, 2
Example
:CALIBRATE:SBOCANCEL:CHANNEL1:
SCHANNEL1 ON
:CALIBRATE:SBOCANCEL:CHANNEL1:
SCHANNEL1?
-> :CALIBRATE:SBOCANCEL:CHANNEL1:
SCHANNEL1 1
Description This command is valid when any of the following
modules is installed.
720254 (4CH 1M16)
:CALibrate:SBOCancel:EXECute (Strain
Balance & Offset Cancel)
Function Executes the strain balancing of all channels on
strain modules or the DC offset canceling of all
channels on voltage measurement modules.
Syntax
:CALibrate:SBOCancel:EXECute
Example
:CALIBRATE:SBOCANCEL:EXECUTE
Description This command is valid when any of the following
modules is installed.
701270(STRAIN_NDIS),
701271(STRAIN_DSUB), 720211 (HS100M12),
720250 (HS10M12), 720254 (4CH 1M16)

Other manuals for YOKOGAWA DL350