Chapter 3 Features and Functions
Burst Mode
3
Waveform Frequency
The
waveform frequency
defines the repetition rate of the burst waveform
in the triggered and external gated modes. In the triggered mode, the
number of cycles specified by the burst count is output at the waveform
frequency. In the external gated mode, the waveform frequency is output
when the external gate signal is true.
Keep in mind that the waveform frequency is different than the “burst
period” which specifies the interval between bursts (triggered mode only).
Waveform frequency: 1
µ
Hz to 20 MHz
(limited to 200 kHz for ramps
and
6 MHz for arbitrary waveforms).
The default waveform frequency
is 1 kHz
.
(For an internally triggered burst waveform, the minimum
frequency is 2.001 mHz.) You can select a sine, square, ramp, pulse,
or arbitrary waveform (noise is allowed only in the gated burst mode
and dc is not allowed).
For sine and square waveforms, frequencies above 6 MHz are allowed
only with an “infinite” burst count.
Front-Panel Operation: To set the waveform frequency, press the
Freq softkey for the selected function. Then use the knob or numeric
keypad to enter the desired frequency.
Remote Interface Operation:
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals