EasyManuals Logo
Home>Agilent Technologies>Inverter>33220A

Agilent Technologies 33220A User Manual

Agilent Technologies 33220A
363 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
94
Chapter 3 Features and Functions
Pulse Width Modulation (PWM)
3
Pulse Waveform
Pulse is the only waveform shape supported for PWM.
Front-Panel Operation:Press .
Remote Interface Operation:
FUNCtion {PULSe}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Pulse Period
The range for the pulse period is 200 ns to 2000 s. The default is 1 ms.
Front-Panel Operation: After selecting the pulse function, press the
Freq softkey again to toggle to the Period softkey. Then use the knob
or numeric keypad to enter the desired pulse period.
Remote Interface Operation:
PULSe:PERiod {<seconds>|MINimum|MAXimum}

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals