Chapter 3 Features and Functions
Amplitude Modulation (AM)
3
Carrier Waveform Shape
AM carrier shape: Sine, Square, Ramp, or Arbitrary waveform.
The default is Sine. You cannot use pulse, noise, or dc as the
carrier waveform.
Front-Panel Operation: Press any of the front-panel function keys
except or . For arbitrary waveforms, press and then
choose the Select Wform softkey to select the active waveform.
Remote Interface Operation:
FUNCtion {SINusoid|SQUare|RAMP|USER}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Carrier Frequency
The maximum carrier frequency depends on the function selected as
shown below. The default is 1 kHz for all functions.
Front-Panel Operation: To set the carrier frequency, press the Freq
softkey for the selected function. Then use the knob or numeric
keypad to enter the desired frequency.
Remote Interface Operation:
FREQuency {<frequency>|MINimum|MAXimum}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.
Function Minimum Frequency Maximum Frequency
Sine
Square
Ramp
Arbs
1 µHz
1 µHz
1 µHz
1 µHz
20 MHz
20 MHz
200 kHz
6 MHz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals