Chapter 3 Features and Functions
Amplitude Modulation (AM)
3
Modulation Depth
The modulation depth is expressed as a percentage and represents the
extent of the amplitude variation. At 0% depth, the output amplitude is
half of the selected value. At 100% depth, the output amplitude equals
the selected value.
Modulation depth: 0% to 120%. The default is 100%.
Note that even at greater than 100% depth, the function generator
will not exceed ±5V peak on the output (into a 50 load).
•If you select the External modulating source, the carrier waveform
is modulated with an external waveform. The modulation depth is
controlled by the ±5V signal level present on the rear-panel
Modulation In connector. For example, if you have set the modulation
depth to 100%, then when the modulating signal is at +5 volts,
the output will be at the maximum amplitude. When the modulating
signal is at -5 volts, then the output will be at the
minimum
amplitude
.
Front-Panel Operation: After enabling AM, press the AM Depth
softkey. Then use the knob or numeric keypad to enter the depth.
Remote Interface Operation:
AM:DEPTh {<depth in percent>|MINimum|MAXimum}

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals