171
Chapter 4 Remote Interface Reference
Using the APPLy Command
4
Output Frequency
For the frequency parameter of the APPLy command, the output
frequency range depends on the function specified. You can substitute
“MINimum”, “MAXimum”, or “DEFault” in place of a specific value
for the frequency parameter. MIN selects the lowest frequency
allowed for the function specified and MAX selects the highest
frequency allowed. The default frequency is 1 kHz for all functions.
Limits Due to Function: The frequency limits are determined by the
function specified in the APPLy command. The APPLy command
always sets both the function and the frequency, and the specified
frequency must be appropriate for the function. For example, the
command APPL:RAMP 20 MHz results in a “Data out of range” error.
The frequency is set to 200 kHz, which is the maximum for a ramp
waveform.
Function Minimum Frequency Maximum Frequency
Sine
Square
Ramp
Pulse
Noise, DC
Arbs
1 µHz
1 µHz
1 µHz
500 µHz
Not Applicable
1 µHz
20 MHz
20 MHz
200 kHz
5 MHz
Not Applicable
6 MHz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals