239
Chapter 4 Remote Interface Reference
Arbitrary Waveform Commands
4
FORMat:BORDer {NORMal|SWAPped}
FORMat:BORDer?
Used for binary block transfers only. Select the byte order for binary
transfers in the block mode using the DATA:DAC command. The default
is NORM. The :BORD? query returns “NORM” or “SWAP”.
•In NORM byte order (default), the most-significant byte (MSB) of
each data point is assumed first.
•In SWAP byte order, the least-significant byte (LSB) of each data
point is assumed first. Most computers use the “swapped” byte order.
The function generator represents binary data as signed 16-bit
integers, which are sent as two bytes. Therefore, each waveform data
point requires 16 bits, which must be transferred as two bytes on the
function generator’s interfaces.
DATA:COPY <destination arb name> [,VOLATILE]
Copy the waveform from volatile memory to the specified name in
non-volatile memory. The source for the copy is always “volatile”.
You cannot copy from any other source and you cannot copy to “volatile”.
The arb name may contain up to 12 characters. The first character
must be a letter (A-Z), but the remaining characters can be numbers
(0-9) or the underscore character (“ _ ”). Blank spaces are not allowed.
If you specify a name with more than 12 characters, a “Program
mnemonic too long” error is generated.
The VOLATILE parameter is optional and can be omitted. Note that
the keyword “VOLATILE” does not have a short form.
The following built-in waveform names are reserved and cannot be
used with the DATA:COPY command: “EXP_RISE”, “EXP_FALL”,
“NEG_RAMP”, “SINC”, and “CARDIAC”. If you specify one of the
built-in waveforms, a “Cannot overwrite a built-in waveform” error is
generated.
The function generator does not distinguish between upper- and
lower-case letters. Therefore, ARB_1 and arb_1 are the same name.
All characters are converted to upper case.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33220A and is the answer not in the manual?

Agilent Technologies 33220A Specifications

General IconGeneral
BrandAgilent Technologies
Model33220A
CategoryInverter
LanguageEnglish

Related product manuals