615
Ohms 10 µA source failed This test configures the 10 V range with
the internal internal 10 M 100:1 divider A4U102 connected across the
input. The 10 µA current source is connected. A 20 ms ADC
measurement is performed and the result is checked against a limit
of 7.5 V ± 3 V.
616 DC current sense failed This test configures the 1 A dc rage and
function. A 20 ms ADC measurement is performed and the result is
checked against a limit of 0 A ± 5 A. This test confirms that the dc
current sense path is functional.
617
Ohms 100 µA source failed This test configures the 10 V range
with the internal 10 M 100:1 divider A4U102 connected across the
input. The 100 µA current source is connected. The compliance limit
of the current source is measured. A 20 ms ADC measurement is
performed and the result is checked against a limit of 5 V ± 1 V.
618 DC high voltage attenuator This test configures to the 300 Vdc
range. the 500 nA ohms current source is connected to produce a
nominal 5 V signal. A 20 ms ADC measurement is performed and the
result is checked against a limit of –10 mV to 70 mV at the output of
the rms-to-dc converter.
619 Ohms 1 mA source failed This test configures the 10 V range with
the internal 1 0 M 100:1 divider A4U102 connected across the input.
The 1 mA current source is connected. A 20 ms ADC measurement is
performed and the result is checked against a limit of 7 V ± 3.5 V.
620 AC rms zero failed This test configures to the 100 mV ac range
with the ac input grounded through A4K103. The internal residual
noise of the ac section is measured and checked against a limit of
–10 mV to 70 mV at the output of the rms-to-dc converter.
621 AC rms full scale failed This test configures for the 100 mV ac
range. The 1 mA ohms current source is switched on the charge the
ac input capacitor A4C301. This produces a pulse on the output of the
rms-to-dc converter which is sampled 100 ms after the current is
applied. A 20 ms A/D measurement is performed and checked against
a limit of 10 V ± 8.5 V into the ADC.
Chapter 6 Service
Self-Test Procedures
170

Table of Contents

Other manuals for Agilent Technologies 34970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34970A and is the answer not in the manual?

Agilent Technologies 34970A Specifications

General IconGeneral
BrandAgilent Technologies
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals