(Optional)
34902A Relay Contact Resistance Verification
This optional procedure uses an external DMM to make 4-wire ohms
measurements across the relay contacts on the 34902A. The measured
resistance is the series resistance of the two relay contacts (both
contacts are in the same relay).
Note: Be sure to read “Plug-in Module Test Considerations on page 87.
Tests 1 - 4: See the diagram on page 97 for the required connections for each test
(be sure to probe the components at the indicated location). For these
measurements, the 34902A is not installed in the 34970A. Record the
4-wire ohms measurements from the external DMM in the table below.
Note: The connections to the external DMM are different for each of
Tests 1, 2, and 3. Be sure to verify the connections shown in the table
below for each of the three tests.
External DMM Ohmmeter Connections
Test
#
HI LO
HI
Sense
LO
Sense
Measured
Value
1 L300 J101, C12 L300 J101, C12 __________ Ohms
2 L301 J101, C14 L301 J101, C14 __________ Ohms
3 L302 J101, C15 L302 J101, C15 __________ Ohms
4 Add (Test 2 + Test 3) __________ Ohms
Chapter 4 Calibration Procedures
Relay Verification
96

Table of Contents

Other manuals for Agilent Technologies 34970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34970A and is the answer not in the manual?

Agilent Technologies 34970A Specifications

General IconGeneral
BrandAgilent Technologies
Model34970A
CategorySwitch
LanguageEnglish

Related product manuals