Continued from previous page...
1 Select each function and range in the order shown in the table below.
Before executing each test, you must press to enable reading
monitoring on the selected channel (or use the ROUTe:MON command
from the remote interface). Compare measurement results to the
appropriate test limits shown in the table (see page 74).
Input
Channel 221
Function
[1]
Range
Quick
Check
Error from Nominal
24 hour 90 day 1 year
Open
Open
Open
DC Current 10 mA
100 mA
1 A
Q
± 1 µA
± 4 µA
± 60 µA
± 2
µA
± 5 µA
± 100 µA
± 2 µA
± 5 µA
± 100 µA
Input
Channel 209
Function
[1]
Range
Quick
Check
Error from Nominal
24 hour 90 day 1 year
Short
Short
Short
Short
Short
DC Volts 100 mV
1 V
10 V
100 V
300 V
Q
± 3.5
µV
± 6 µV
± 40 µV
± 600 µV
± 6 mV
± 4 µV
± 7 µV
± 50 µV
± 600 µV
± 9 mV
± 4 µV
± 7 µV
± 50 µV
± 600 µV
± 9 mV
Short
Short
Short
Short
Short
Short
Short
2-Wire Ohms
[2]
and
4-Wire Ohms
100
Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
Q
± 3.5 mΩ
±
6 mΩ
±
50 mΩ
±
500 mΩ
±
10 Ω
±
100 Ω
±
10 kΩ
±
4 mΩ
±
10 mΩ
±
100 mΩ
±
1 Ω
±
10 Ω
±
100 Ω
±
10 kΩ
±
4 mΩ
±
10 mΩ
±
100 mΩ
±
1 Ω
±
10 Ω
±
100 Ω
±
10 kΩ
[1] Select 6
1
⁄
2
digit resolution.
[2] For 2-wire ohms, an additional 4
Ω of error must be added.
Q: Quick performance verification test points.
Note: Zero offset calibration using a multifunction calibrator is
NOT recommended. The calibrator and cabling offset can be large
and unstable causing poor offset calibration of the internal DMM.
Chapter 4 Calibration Procedures
Internal DMM Verification Tests
76