EasyManuals Logo
Home>Agilent Technologies>Switch>34970A

Agilent Technologies 34970A Service Guide

Agilent Technologies 34970A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
AC CURRENT Gain Verification Test
Configuration: AC Current
LF 3 HZ:SLOW (in the Advanced menu)
1 Make sure you have read “Test Considerations” on page 71.
2 Select Channel 221, set the AC CURRENT function and the 3 Hz
input filter. With the slow filter selected, each measurement takes
7 seconds to complete. Before executing each test, you must press
to enable reading monitoring on the selected channel (or use the
ROUTe:MON command from the remote interface).
3 Select each range in the order shown below. Compare measurement
results to the appropriate test limits shown in the table. (Be certain
to allow for appropriate source settling.)
Input
Current Frequency
Range
Quick
check
Error from Nominal
24 hour 90 day 1 year
10 mA
[1]
100 mA
[1]
10 mA
1A
[1]
1 kHz
1 kHz
1 kHz
1 kHz
10 mA
100 mA
1 A
1A
Q
± 14 µA
± 600 µA
± 1.41 mA
± 1.4 mA
± 14 µA
± 600 µA
± 1.41 mA
± 1.4 mA
± 14 µA
± 600 µA
± 1.41 mA
± 1.4 mA
[1] Verify only, no adjustment.
Frequency Gain Verification Test
Configuration: Frequency
6
1
2
digits
1 Make sure you have read “Test Considerations” on page 71.
2
Select Channel 210, select the FREQUENCY function and set 6
1
2
digits.
3 Select each range in the order shown below. Compare measurement
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling.)
Input
Voltage Frequency
Range
Quick
Check
Error from Nominal
24 hour 90 day 1 year
10 mV
[1]
1 V
100 Hz
100 kHz
100 mV
1 V
Q
± 0.06 Hz
± 6 Hz
± 0 .1 Hz
± 10 Hz
± 0 .1 Hz
± 10 Hz
[1] Verify only, No adjustment. For this test, isolate the calibrator’s output from earth ground.
Q: Quick performance verification test points.
4
Chapter 4 Calibration Procedures
Internal DMM Verification Tests
79

Table of Contents

Other manuals for Agilent Technologies 34970A