... Continued from previous page
External DMM Ohmmeter Connections
Test
#
Channel
Closed*
HI LO
HI
Sense
LO
Sense
Measured Value
Test
Limit
Relay
Measured
18 Ch 8 P2 P2 P1 P1 __________ Ohms
2.00Ω
K408
19 Ch 9 P2 P2 P1 P1 __________ Ohms
2.00Ω
K409
20 Ch 10 P2 P2 P1 P1 __________ Ohms
2.00Ω
K410
21 Ch 11 P2 P2 P1 P1 __________ Ohms
2.00Ω
K411
22 Ch 12 P2 P2 P1 P1 __________ Ohms
2.00Ω
K412
23 Ch 13 P2 P2 P1 P1 __________ Ohms
2.00Ω
K413
24 Ch 14 P2 P2 P1 P1 __________ Ohms
2.00Ω
K414
25 Ch 15 P2 P2 P1 P1 __________ Ohms
2.00Ω
K415
26 Ch 16 P2 P2 P1 P1 __________ Ohms
2.00Ω
K416
27 Ch 17 P2 P2 P1 P1 __________ Ohms
2.00Ω
K417
28 Ch 18 P2 P2 P1 P1 __________ Ohms
2.00Ω
K418
29 Ch 19 P2 P2 P1 P1 __________ Ohms
2.00Ω
K419
30 Ch 21 P4 P4 P5 P5 __________ Ohms — —
31 Subtract (Test 30 – Test 4) __________ Ohms
2.00Ω
K522
32 Ch 22 P4 P4 P5 P5 __________ Ohms — —
33 Subtract (Test 32 – Test 5) __________ Ohms
2.00Ω
K522
* Only the channel currently under test should be closed at one time. All other channels should be open.
Module Reference
4
Chapter 4 Calibration Procedures
Relay Verification
93