DTA 100 C Device settings
822-129-8 35 / 90
8.4 Overwrite measurement logs
The device can store up to ten measurement logs. When the measurement log memory is
full, old measurement logs must be deleted before you can save new ones. Under Clear
meas. logs you can define how to proceed when the measurement log memory is full and a
measurement is to be performed.
Main menu > Device settings > Clear meas. logs
1. Choose one of the following settings by pressing the arrow keys.
yes: When you start the next measurement log, the oldest measurement log will be
automatically deleted.
no: Before starting the next measurement log, a saved measurement log must be
deleted manually.
Further information: Chapter Display of measurement results (on page 64)
2. Click Save to confirm the entry.
8.5 Configure the IEC 60156:2018 settings
According to the standard IEC 60156:2018, its Annex A and the standards based on
IEC 60156:2018, it is possible to adjust the hold time prior to the first measurement, various
stirrer settings and electrode shapes. The settings configured under this menu item are
applied automatically for the following standards:
IEC 60156:2018
IEC 60156:2018 Annex A
AS 1767.2.1
BS EN 60156
CEI EN 60156
SEV EN 60156
UNE EN 60156
NF EN 60156
SABS EN 60156
VDE 0370-5:1996