Device settings DTA 100 C
5. Click Save to confirm the entry.
A window appears on the display with an input field for the distance between electrodes.
6. Use the number keys on the membrane keypad to enter the desired distance between
electrodes.
A distance between electrodes of between 0.5 and 10 mm can be entered. The entered
distance between electrodes is displayed in the measurement log.
7. Click Save to confirm the entry.
8.8 Set date
Main menu > Device settings > Date
1. Select the position by pressing the arrow keys.
2. Change the date with the membrane keypad.
3. Click Save to confirm the entry.