DTA 100 C User-defined measurement
822-129-8 61 / 90
11.2 Running a user-defined measurement
Main menu > User-defined measurement
1. In the main menu, select the menu item User-defined measurement.
2. Select the menu item Next to confirm the selection.
3. In the User-defined measurement menu, select a saved measurement, followed by the
menu item Next.
4. Select the menu item run, followed by the menu item Next.
A prompt appears on the display to set an electrode gap. The required electrode shape
is displayed pictographically.
5. Check whether the correct electrodes have been selected. If the wrong electrodes are
used, replace them.
Further information: Chapter Replacing the electrodes (on page 27)