8.6 Configure the ASTM settings
With the ASTM D standards, it is possible to display the individual breakdown values and set
the hold time before the first measurement. The settings configured under this menu item are
applied automatically for the following standards:
ASTM D1816:2012 1 mm
ASTM D1816:2012 2 mm
ASTM D877/D877M:2013 PA
ASTM D877/D877M:2013 PB
Main menu > Device settings > ASTM settings
8.6.1 Display individual breakdown values
Main menu > Device settings > ... > Display indiv. breakdown values
1. In the main menu, select the menu item Device settings.
2. Select the menu item Next to confirm the selection.
3. In the Device settings menu select the menu item ASTM settings and then menu item
Next.
4. Select menu item Display indiv. breakdown values.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
yes: The individual breakdown values recorded during a series are listed in the
measurement log.
no: The average value of the recorded breakdown values is calculated and
displayed in the measurement log.
7. Click Save to confirm the entry.