DTA 100 C Standardised measurement
822-129-8 45 / 90
2)
Can be selected under Main menu > Device settings > ASTM settings
When testing ester liquids in accordance with ASTM D1816:2012 the hold time before the
first measurement must be at least 30 mins.
3)
Can be deactivated under Main menu > Device settings > IEC 60156:2018 settings
4)
Measurements per filling / number of measurements
The electrode shapes and their dimensions:
:
IEC 60156 Fig. I, ball Ø 12.5 to 13.0
:
IEC 60156 Fig. II, Ø 36.0 / radius 25.0
:
ASTM D877, disc Ø 25.4
9.2 Carry out a standardised measurement
Main menu > Standardised measurement
1. In the main menu, select the menu item Standardised measurement.
2. Select the menu item Next to confirm the selection.