Main menu > Device settings > IEC 60156:2018 settings
8.5.1 Setting the hold time before the first measurement
Main menu > Device settings > IEC 60156:2018 settings > Hold time
1. In the main menu, select the menu item Device settings.
2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the IEC 60156:2018 settings menu item and then
Next.
4. Select Hold time menu item.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
5 min: The hold time is 5 minutes.
15...60 min: The hold time can be set to between 15 and 60 minutes.
7. If you have enabled the 15...60 min setting, select the Next menu item.
8. Enter the desired hold time.
For insulating liquids with a standard viscosity greater than 15 mm
2
/s (40°C), the hold
time prior to the first measurement should be 15-30 minutes.
9. Click Save to confirm the entry.