EasyManuals Logo
Home>Baur>Test Equipment>DTA 100 C

Baur DTA 100 C User Manual

Baur DTA 100 C
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
Device settings DTA 100 C
Main menu > Device settings > IEC 60156:2018 settings
8.5.1 Setting the hold time before the first measurement
Main menu > Device settings > IEC 60156:2018 settings > Hold time
1. In the main menu, select the menu item Device settings.
2. Select the menu item Next to confirm the selection.
3. In the Device settings menu, select the IEC 60156:2018 settings menu item and then
Next.
4. Select Hold time menu item.
5. Select menu item Change.
6. Choose one of the following settings by pressing the arrow keys:
5 min: The hold time is 5 minutes.
15...60 min: The hold time can be set to between 15 and 60 minutes.
7. If you have enabled the 15...60 min setting, select the Next menu item.
8. Enter the desired hold time.
For insulating liquids with a standard viscosity greater than 15 mm
2
/s (40°C), the hold
time prior to the first measurement should be 15-30 minutes.
9. Click Save to confirm the entry.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Baur DTA 100 C and is the answer not in the manual?

Baur DTA 100 C Specifications

General IconGeneral
BrandBaur
ModelDTA 100 C
CategoryTest Equipment
LanguageEnglish

Related product manuals