DTA 100 C User-defined measurement
822-129-8 59 / 90
6. Select the menu item Next to confirm the selection.
7. Select the duration for which the withstand voltage is maintained during the first step.
8. Select the menu item Next to confirm the selection.
9. Select the withstand voltage value and the duration of the withstand voltage for all
subsequent steps. To do this, repeat steps 5 to 8 for each further step.
11.1.11 Selecting which measured values are not to be evaluated
Define which measured values are considered in the measurement result.
Main menu > User-defined measurement > ... > Ignore measurement values
1. Select the menu item by pressing the arrow keys.
2. Select menu item Change.
3. Choose one of the following settings by pressing the arrow keys:
none: The measured values of all measurements are included in the measurement
result.
first: The first measurement is not included in the measurement result.
last: The last measurement is not included in the measurement result.
first & last: The first and last measurements are not included in the measurement
result.
highest result: The maximum measured value is not included in the measurement
result.
4. Click Save to confirm the entry.