EC9830 CO ANALYZER OPERATION MANUAL
1-2 98307600 Rev. C-1
1.1 Specifications
Note
All specifications are referenced to STP (standard
temperature and pressure).
1.1.1 Range
Display: Autoranging 0 to 200 ppm. Resolution = 1 ppt (selectable units and
decimal places).
Analog output: 0-full scale from 0-1 ppm to 0-200 ppm with 0%, 5%, 10%
offset.
Autoranging between 2 user-specified full scale values.
U.S. EPA designated range: Any full scale range between 0-5.0 ppm and
0-100 ppm.
1.1.2 Noise (RMS)
Measurement process: 0.025 ppm or 0.1% of concentration reading,
whichever is greater, with Kalman filter active.
Analog output: 0.025 ppm or 0.1% of analog output full scale, whichever is
greater.
1.1.3 Lower Detectable Limit
Measurement process: Less than 0.05 ppm or 0.2% of concentration reading,
whichever is greater, with Kalman filter active.
Analog output: 0.05 ppm or 0.2% of analog output full scale, whichever is
greater.
1.1.4 Zero Drift
Temperature dependent, 0.01 ppm per °C.
Time dependent, at fixed temperature:
24 hours: Less than 0.1 ppm
30 days: Less than 0.1 ppm
1.1.5 Span Drift
Temperature dependent, 0.05% per °C.
Time dependent, at fixed temperature:
24 hours: 0.5% of reading.
30 days: 0.5% of reading.