EC9830 CO ANALYZER OPERATION MANUAL
2-18 98307600 Rev. C-1
TIME
Set in 24-hour format. Setting the time resets the seconds (internally) to zero for
synchronization with an external clock.
PASSWORD
See section 2.7 below.
ERASE MEMORY
If YES is selected, the following message is displayed:
!THIS WILL ERASE SYSTEM GAINS!
!!!ARE YOU SURE: NO
The word NO is highlighted in this warning. Scrolling to YES and pressing
<Enter> will erase the memory in the analyzer.
This feature is provided for service, and for preliminary configuration purposes.
Please do not choose this selection during normal operation.
Caution
If the analyzer memory is erased, all user-
configured parameters will return to their
default values. In addition, all instrument
calibration will be lost, so the analyzer will have
to be fully recalibrated. This feature is provided
for service, and for preliminary configuration
purposes. Please do not choose this selection
during normal operation.
MAIN GAS ID
The ID address of the analyzer when Multidrop RS232 communications is used.
2.5.4 Measurement Menu
MEASUREMENT MENU
UNIT SELECTION : uG/M3
CONVERSION TEMP : 0 DEG C*
DECIMAL PLACES : 3
AVERAGE PERIOD : 1 MINUTE
FILTER TYPE : KALMAN
NOISE : 24.234 PPB
ZERO OFFSET : 0.00 PPB
Figure 2-10. Measurement Menu