EasyManuals Logo
Home>Ecotech>Measuring Instruments>A Series

Ecotech A Series User Manual

Default Icon
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
EC9830 CO ANALYZER OPERATION MANUAL
2-20 98307600 Rev. C-1
FILTER TYPE
Sets the time constant of the digital filter. Choices are NO FILTER, 300
SECONDS, 90 SECONDS, 60 SECONDS, 30 SECONDS, 10 SECONDS or KALMAN
(adaptive).
Note
The Kalman filter is the factory default setting and
must be used when using the instrument as a U.S.
EPA equivalent method.
NOISE
The standard deviation of the concentration. The manner in which this is done is
as follows: 1. Take a concentration value once every two minutes; 2. Store 25 of
these samples in a first in, last out buffer; 3. Every two minutes, calculate the
standard deviation of the current 25 samples. This is a microprocessor-generated
field and cannot be set by the operator.
Note
This reading is only valid if zero air or a steady
concentration of span gas has been fed to the
analyzer for at least one hour.
ZERO OFFSET
ZERO calibration correction factor. User can manually set the offset between
± 10.00 PPB.
2.5.5 Calibration Menu
The CALIBRATION MENU contains entries used to choose calibration gases or to
perform automatic zero-span cycles. The choice of TIMED or MANUAL calibration
displays a slightly different screen. TIMED calibration generates a zero/span check
that occurs at a chosen interval without operator intervention. MANUAL calibration
allows for operator-controlled calibration. Only one choice, TIMED or MANUAL,
applies at any given time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ecotech A Series and is the answer not in the manual?

Ecotech A Series Specifications

General IconGeneral
BrandEcotech
ModelA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals