CHAPTER 1, DESCRIPTION
98307600 Rev. C-1 1-3
1.1.6 Lag Time
Less than 20 seconds.
1.1.7 Rise/Fall Time, 95% of Final Value
Less than 40 seconds (1 slpm flow) with Kalman filter active.
1.1.8 Linearity Error
±1% of full scale (0 to 50 ppm); ±2% of full scale (0 to 200 ppm), from best
straight-line fit.
1.1.9 Precision
0.1 ppm or 1% of reading, whichever is greater.
1.1.10 Sample Flow Rate
1 slpm.
1.1.11 Sample Pressure Dependence
A 5% change in pressure produces less than 1% change in reading.
1.1.12 Temperature Range
5° to 40° C (41° to 104° F).
U.S. EPA designated range: 15° to 35° C.
Eignungsgeprüft range: 5° to 40° C.
1.1.13 Power
99 to 132 VAC; 198 to 264 VAC; 47 to 63 Hz.
U.S. EPA designated range: 105 to 125 VAC, 60 Hz, or 200 to 240 VAC, 50
Hz.
1.1.14 Weight
20.9 kg (46 lb).
1.1.15 Analog Output
Menu selectable current output of 0-20, 2-20, and 4-20 mA.
The 9830B requires an optional 50 I/O board to give voltage output. Jumper
selectable voltage output of 100 mv, 1 volt, 5 volts, and 10 volts, with menu
selectable zero offset of 0%, 5%, or 10%.

Table of Contents