Revision L MAC™ 3500 Resting ECG Analysis System 5-37
2021337-036
Maintenance: Functional Checkout Procedures
9. Fasteners replaced and secured?
10. All harnesses’ and internal wiring has been secured?
Checkout Procedures
Perform the applicable checkout procedure.
Operational Checks
1. Power-up self-test passed?
2. Simulated recorded rhythm strip successful?
3. Simulated recorded ECG successful?
4. Simulated ECG stored on media successful?
5. Simulated ECG data transmitted successfully to MUSE?
6. Exercise device successfully communicates with MAC5500/5000?
7. Blood pressure monitor successfully communicates with
MAC5500/5000?
8. Barcode reader successfully communicates with MAC5500/5000?
9. Magnetic card reader successfully communicates with MAC5500/5000?
10. Trolley casters lock successfully?
Diagnostic Tests
11. Battery Status Test meets Battery Current expectation?
12. Display diagnostic test successful?
13. Speaker test successful?
14. Keyboard test successful?
15. Writer diagnostic tests successful?
16. Read/Write Floppy Diagnostic Test successful? (Accurate test does not have to
be performed
)?
17. External/Internal Modem Test successful?
18. SD Card Test successful?
19. Ethernet Module Test successful?