24
APT4000PH
Operation: Keypad
Start, end calibration
Start, end configuration
Select digit position
(selected position flashes)
Edit digit
Calibration:
Continue in program sequence
Configuration: Confirm entries,
next configuration step
Measuring mode: Display output current
Cal Info, display of asymmetry potential
and slope
Error Info, display last error message
Start GainCheck device self-test
+
25
Sensoface provides information on the electrode condi-
tion. The asymmetry potential (zero), slope and
response time during calibration are evaluated. The
three Sensoface indicators provide the user with infor-
mation about wear and required maintenance of the
sensor.
Note: When measuring with DURAFET electrodes,
Sensocheck is not active.
GainCheck device self test
A display test is carried out, the software version is displayed
and the memory and measured value transfer are checked.
Start GainCheck device self-test:
Automatic device self-test
The automatic device self-test checks the memory and meas-
ured-value transfer. It runs automatically in the background at
fixed intervals.
+
Safety functions
Sensocheck, Sensoface sensor monitoring
Sensocheck continuously monitors the sensor and leads.
Sensocheck can be switched off (Configuration, Pg 51).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell APT4000PH and is the answer not in the manual?

Honeywell APT4000PH Specifications

General IconGeneral
BrandHoneywell
ModelAPT4000PH
CategoryMeasuring Instruments
LanguageEnglish