46
APT4000PH
Configuration
Temperature compensation
Temp detection for meas/cal. TC process medium
Temp during measurement
Temp during calibration
Enter TC process medium
End:
Press CONF, then ENTER
Temperature
compensation
tc.
Code
Display
Menu group
Select menu item
47
AUT
(MAN)
Code
Display
Action
Choices
tc.
Select temp detection during meas-
urement (Auto/MAN)
AUTO: Temp detection with temper-
ature probe
MAN: Manual temperature input
Select with key,
proceed with ENTER
Only with manual temp detection
selected (MAN)
Enter temperature.
Select position with key,
edit number with key,
proceed with ENTER
25 °C
(XXX.X)
Select temp detection during
calibration (Auto/MAN)
Select with key,
proceed with ENTER
AUT
(MAN)
Only with manual temp detection
selected (MAN)
Enter temperature.
Select position with key,
edit number with key,
proceed with ENTER
For pH measurement only:
Enter temperature compensation of
process medium
Select position with key,
edit number with key,
proceed with ENTER
00.00
%/K
(-19.99
to 19.99
%/K)
25 °C