APT4000PH
Configuration
Calibration mode
Calibration mode
Cal timer interval
Code
Display
Menu group
Select menu item
Calibration mode
CA.
End:
Press CONF, then ENTER
49
-O4-BUF
(-01-BUF/
-02-BUF/
-03-BUF/
-04-BUF/
-05-BUF/
-06-BUF/
-07-BUF/
MAN/
DAT)
Code
Display
Action
Choices
CA.
For pH measurement only:
Select calibration mode
BUF: Calibration with Calimatic auto-
matic buffer selection.
To do so, you must select your buffer
set:
-01- BUF: Mettler-Toledo
-02-BUF: Merck Titrisols, Riedel Fixanals
-03-BUF: Ciba (94)
-04-BUF: NIST technical buffers
-05-BUF: NIST standard buffers
-06-BUF: HACH buffers
-07-BUF: WTW technical buffers
MAN: Calibration with
manual buffer entry
DAT: Entry of asymmetry potential
and slope of premeasured electrodes
Select with key,
proceed with ENTER
Enter calibration interval:
Entry of time interval within which
the analyzer is to be calibrated.
With a time interval of 0000 hrs the
calibration timer is not active.
Select position with key,
edit number with key,
proceed with ENTER
0000 h
(0000 to
9999 h)