APT4000PH
Overview of configuration steps
Time constant of output filter
22 mA signal in the case of error
Signal behavior during Hold
Enter fixed value
out2
Output 2
o2.
Select temperature unit
Select temperature probe
Select current range
Enter current beginning
Enter current end
Time constant of output filter
22 mA signal in the case of temp error
Signal behavior during Hold
Enter fixed value
tc.
Temperature compensation
tc.
Temp detection during meas
Temp detection during cal
Enter TC process medium
CAL
Calibration mode
CA.
Select calibration mode
Enter cal timer interval
Menu
Code
out1
Output 1
o1.
Select measured variable
Select electrode type
Select current range
Enter current beginning
Enter current end
Selection / Default
pH / ORP
GLAS EL / FEt EL
0-20 mA / 4-20 mA
XX.XX pH / XXXX mV (00.00 pH)
XX.XX pH / XXXX mV (14.00 pH)
0 .... 120 SEC (0 SEC )
ON / OFF
LAST / Fix
000.0 ... 021.0 mA (021.0 mA)
°C / °F
Pt100/Pt1000/NTC30/ NTC8.55
0-20 mA / 4-20 mA
XXX.X (000.0 °C )
XXX.X (100.0 °C )
0 ... 120 (0 SEC)
ON / OFF
LAST / Fix
000.0 ... 021.0 mA (021.0 mA)
Auto/man (man: XXX.X°C (025.0°C))
Auto/man (man: XXX.X°C (025.0°C))
-19.99 ... +19.99%/K (00.00%/K)
-04-BUF / MAN / DAT
0000 ... 9999 h (0000 h)
31
Menu
Code
Selection / Default
ALrt
Alarm settings
AL.
Select Sensocheck
Enter alarm delay
LED in Hold mode
rLAY
Relay 1/2: Limit values, controller
rL.
Select limit function / controller
Select contact function
L1.
Select contact response
Enter switching point
Enter hysteresis
Enter delay
Select contact function
L2.
Enter controller setpoint
Ct.
Enter neutral zone
(P) Controller gain K
P
(I) Reset time T
R
(D) Rate time T
D
Controller
PLC: Pulse length
PFC: Pulse frequency
Select Hold behavior
PrbE
Rinsing and cleaning probes
Pb.
Select cleaning / calibration probe
Rinsing interval
rinse
Rinse duration
Contact response
Cleaning interval
EASYCLN
Select contact response
Enter switching point
Enter hysteresis
Enter delay
Calibration interval
Lock cleaning / calibration interval
ON /
OFF
0000 ... 0600 SEC (0010 SEC)
ON / OFF
LiMIT / CtROL
Lo / Hi
N/O / N/C
XX.XX pH / XXXX mV
(00.00 pH)
XX.XX pH / XXXX mV
(00.50 pH)
0000 ... 9999 SEC (0010 SEC)
Lo / Hi
XX.XX pH / XXXX mV
(07.00 pH)
XX.XX pH / XXXX mV
(01.00 pH)
0010 ... 9999 % (
100 %
)
0000 ... 9999 SEC (0000 SEC)
0000 ... 9999 SEC (0000 SEC)
PLC / PFC
0001 ... 0600 SEC (0010 SEC)
0001 ... 0180 /min (0060 /min)
Y LAST / Y Off
EASYCLN / rinse
000.0 ... 999.9 h (
000.0 h
)
0000 ... 1999 SEC (0060 SEC)
N/O /
N/C
000.0 ... 999.9 h (
000.0 h
)
N/O / N/C
XX.XX pH / XXXX mV
(
14.00 pH
)
XX.XX pH / XXXX mV
(00.50 pH)
0000 ... 9999 SEC (
0010 SEC
)
000.0 ... 999.9 h (000.0 h)
ON / OFF