EasyManuals Logo
Home>Keithley>Multimeter>DMM7510

Keithley DMM7510 Calibration Manual

Keithley DMM7510
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Model DMM7510
7½ Digit Graphical Sampling Multimeter Calibration Manual Section 2:
Performance verification
DMM7510-905-01 Rev. A / April 2015 2-15
Verify AC voltage accuracy for the 700 V range
Use shielded low capacitance cabling. For the 700 V range, avoid cable capacitances of >150 ρF.
Excessive capacitance may result in additional load regulation uncertainties and could cause the
calibrator output to open (go into standby).
To verify AC voltage accuracy for the 700 V range:
1. Put calibrator in Standby.
2. Connect the Model DMM7510 HI and LO INPUT connectors to the calibrator as shown in the
following figure.
3. For 700 V at 50 kHz and 100 kHz outputs, connect the calibrator to the Fluke 5725A amplifier.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
BrandKeithley
ModelDMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals