2: Performance verification Model DMM7510 7½ Digit Graphical Sampling Multimeter
2-14 DMM7510-905-01 Rev. A / April 2015
2. On the Model DMM7510, press the FUNCTION key and select AC voltage.
3. Press the HOME key.
4. Select Range and select 100 mV.
5. Press the MENU key.
6. Select Settings.
7. Ensure that Detector Bandwidth is set to 30 Hz.
AC voltage is specified for the detector bandwidth setting of 3 Hz. 3 Hz measures accurately for input
signals from 3 Hz to 300 kHz, with reading rates ≈0.5 readings/s. To improve verification throughput
to ≈3.3 readings/s, set detector bandwidth to 30 Hz for frequencies of 30 Hz to 300 kHz. To verify
frequencies 1 kHz and higher, set the detector bandwidth to 300 Hz for faster ≈55 readings/s
throughput.
8. Source AC voltages for each of the frequencies listed in the Verify the AC voltage 100 mV range
(on page 2-14) table.
9. Repeat these steps for each range and frequency listed in the tables below. For each voltage
setting, be sure that the reading is within low and high limits.
Verify the AC voltage 100 mV range
Input Frequency Lower limit Upper limit
Verify the AC voltage 1 V range
Input Frequency Lower limit Upper limit
1 1.0E+03 9.991000E-01 1.000900E+00
Verify the AC voltage 10 V range
Input Frequency Lower limit Upper limit
10 1.0E+03 9.991000E+00 1.000900E+01
10 5.0E+04 9.981000E+00 1.001900E+01
Verify the AC voltage 100 V range
Input Frequency Lower limit Upper limit