Doc# 0296-0018 • REV R (July 2017) Page 19 of 61
FIG. 3-11 INSTALLATION OF MOUNTING BRACKET FOR AXIS A & B
Step 2. While holding the reference accelerometer steady with the spanner wrench, place the mounng
bracket (Prt. #0205- 0043-01) over the top of the reference accelerometer.
WARNING ALWAYS USE THE SPANNER WRENCH WHEN INSTALLING, REMOVING OR ADJUSTING THE
TRANSDUCERS OR MOUNTING BRACKET ON THE REFERENCE ACCELEROMETER.
Step 3. While sll holding the spanner wrench in place, use the allen wrench provided with the bracket
kit to screw the mounng bracket socket head screw into the reference accelerometer. (See Fig. 3-11)
Tighten the screw so that the bracket is snug against the reference accelerometer. If you want use a
torque wrench and torque to 18 inch pounds. Do not over torque the mounng screw. Overghtening
the mounng screw can cause bad readings.
Step 4. Install the connecng cable (Prt. # 0214-0088-01) between the Transducer under test (TUT) and
the input connector on the top panel. (See Fig. 3-12 & 3-13)
Triaxial Mounting Bracket
Mounting Bracket Screw
Spanner Wrench
Prt. #0228-072-01
Reference Accelerometer