AS3000™ Service Manual 0070-10-0683 5 - 3
Calibration Calibration Warnings, Precautions, and Notes
5.2.3 Notes
NOTE: Only bacterial filters with a low flow resistance must be
connected to the patient module and/or the patient
connection.
NOTE: Use surgical gloves whenever touching or disassembling
valves or other internal components of the Breathing
System.
NOTE: Ensure that the gas supply of the machine always complies
with the technical specification.
NOTE: The APL Valve and PAW gauge numerics are for reference
only. Calibrated patient airway pressure is available on the
ventilator screen.
NOTE: If the machine should show faults during the initial
calibration or testing, the machine should not be operated
until the fault has been repaired by a qualified service
technician.
NOTE: After servicing, functional, sensor and system tests must be
carried out before clinical use.
NOTE: To accommodate additional monitors and other equipment
the AS3000 offers up to two vertical mounting tracks. Use
of unauthorized mounting accessories is not recommended.
NOTE: Always secure any equipment placed on the AS3000’s top
shelf.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals