AS3000™ Service Manual 0070-10-0683 6 - 3
Periodic Maintenance Replacement of Consumable Parts
6.5 Replacement of Consumable Parts
Parts are replaced at multiple intervals from the date of installation.
CONSUMABLE PARTS AMOUNT 12 MONTH PM 36 MONTH PM PART NUMBER
1O
2
Sensor 1 check check 0600-00-0149
2 Battery Assembly SLA-5AH24V 1 check replace 0997-00-0619
3 Drive Gas Tube 1 check replace 0004-00-0098
4 Bellows Assembly 1 replace replace 115-005968-00
5 Scavenger Hose 1 check replace 0004-00-0095
6 I/E Valve gasket 2 replace replace 0354-00-0192
7 O-ring, 6 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0188
8 O-ring, 4 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0189
9 Bag Arm retainer ring 1 check check 0354-00-0039
10 Disc, Ceramic I/E Valve 2 check replace 0354-00-0182
11 Gasket, Bellows canister base 1 replace replace 0354-00-0191
12 Absorber canister Seal
3 check replace 0354-00-0183 (original)
1
2
check
check
replace
replace
047-002850-00 (revised)
047-000252-00 (revised)
13 O-ring O
2
Interface / O
2
Plug 2 replace replace 0354-00-0187
14 O-ring Vaporizer mount 14x2.65 4 replace replace 0354-00-0193
15 Cylinder Washer, Black 3 replace replace 0348-00-0185
16 Absorber assembly 1 clean clean 0997-00-0629 (original)
115-005724-00 (revised)
17 AGSS filter 1 clean replace 0378-00-0074
18 Absorber Gasket 2 clean replace 0354-00-0199 (original)
047-002849-00 (revised)
19 Check Valve
(Applicable to original model P/N 0997-00-0618
only. See “Check Valve Cleaning” on page 6-4.)
1 clean clean Non-replaceable
20 User Interface cable jackscrews and jackposts 2 Check for Loctite
243 and apply if
necessary. Verify
jackscrews are
secured tightly.
Check for Loctite
243 and apply if
necessary. Verify
jackscrews are
secured tightly.
Non-replaceable

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals