EasyManuals Logo
Home>Mindray>Medical Equipment>Datascope AS 3000

Mindray Datascope AS 3000 User Manual

Mindray Datascope AS 3000
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #128 background imageLoading...
Page #128 background image
AS3000™ Service Manual 0070-10-0683 6 - 1
6.0
Periodic Maintenance
6.1 Maintenance Schedule
The following is a list of activities required for periodic maintenance of the AS3000
Anesthesia System. Physical inspection, replacement of consumables and performance
checks should be periodically performed per the schedule listed below. Certain calibration
adjustments are required only after replacing one or both of the active devices. The
manufacturer is not responsible for component failure or loss resulting from the use of stated
consumables beyond their recommended replacement interval. These are noted in the
Periodic Maintenance Schedule (See section 6.3).
6.2 Periodic Maintenance Consumable Parts Kits
Consumable parts are available in the PM kits listed below.
AS3000, 12-Month PM Kit: 0040-00-0446
AS3000, 36-Month PM Kit: 0040-00-0445
Parts must be replaced at periodic intervals according to the schedule.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals