AS3000™ Service Manual 0070-10-0683 3 - 11
Repair Information Leak Troubleshooting
3.6 Leak Troubleshooting
FIGURE 3-1 Leak Troubleshooting Flow Chart
PASSLeak Test
Manual
Leak Test
PASS
No Leaks
Detected
FAIL FAIL
Manual
Leak Test
Seal the
scavenger port
and repeat the
test
PASS PASS
Check APL
Valve
FAIL FAIL
Check the
Manual bag
and bag arm
Check the
Exhaust valve
Check the
Bellows Dome
Check the
Proportional
Valve
Check the PEEP
valve
Pinch the CGO
Hose and
repeat the Leak
test
PASS
Remove the
Vaporizers
and repeat the
Leak test
PASS
Check the
vaporizers and
the vaporizer
mount
FAIL FAIL
Check the
Flowmeter
module /
CGO port
Remove
sampling
line
Check and if
necessary
exchange the
breathing hoses
Check
the O2
cell
o-ring
Check
the
Absorber
assembly
Verify the I/E Rings
are tight and not
cracked (do not
over tighten!)
Check the
PAW
gauge and
its o-rings

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals