AS3000™ Service Manual 0070-10-0683 3 - 3
Repair Information Special Tools Required
3.4 Special Tools Required
PART NUMBER DESCRIPTION SPECIFICATION
1 0070-00-0683 AS3000 Service Manual NA
2 not applicable Vaporizer Instruction Manual NA
3 not applicable Safety Analyzer Dempsey 430 or equivalent
4 not applicable Digital Volt Meter 3 1/2 digit
5 not applicable Agent (and N
2
O) Analyzer ±0.3 V/V% + 5% of reading
6 0138-00-0012 Test Lung, Adult NA
7 not applicable Digital Pressure Meter BC Biomedical DPM-2301751
NMC Digital Pressure Meter
8 not applicable Central supplied O
2
, N
2
O, AIR Minimum of 35 psi, DISS
connections.
9 not applicable Cylinder gases O2, N
2
O, AIR Full PISS yoke connections.
10 not applicable Hand tools, Allen wrench set Metric
11 0367-00-0080 Tank Wrench NA
12 0103-00-0508 Y-Fitting 15 mm connection
13 0004-00-0076 Respiration Tube (2 required) 0.6 meter silicone, 15 mm
14 0992-00-0139 Breathing Bag 2.3 L silicone
15 0510-00-0020 Krytox Lubricant NA
16 not applicable Vent Tester Certifier-FA PLUS Respiratory
Vent.Tester 4080
17 0997-00-0641 AS3000 Calibration hose NA
18 0103-00-0691 AS3000 test plug NA
19 0040-00-0448 Plug Kit NA
20 0453-00-1216 Regulator Calibration Hose NA