AS3000™ Service Manual 0070-10-0683 2 - 5
Installation Guide Assembly
2. Discard the washers that came with the vaporizer. Use only approved O-rings that come
mounted on the vaporizer mounting ports of the AS3000.
3. Mount a mechanical vaporizer to either side of the AS3000 vaporizer mount by
placing the vaporizer on the ports and locking down the lever.
4. Mount an electronic vaporizer in the same manner. Follow any installation instructions
that come with the specific vaporizer.
2.2.8 High Pressure Hoses
1. Remove the hoses from their packing material.
2. Attach each hose to its mating connector by hand. Do not use any tools.
3. Use extreme care while attaching the EVAC hose to the waste gas scavenger (WGS); as
it is extremely delicate.
4. Attach an EVAC hose to a VAC source connection only if an EVAC source connection is
not available.
2.2.9 Emergency Cylinder(s)
1. Remove the cover from a new O
2
, N
2
O, and AIR cylinder.
2. Mount one at a time onto the rear of the anesthesia machine.
3. Discard the cylinder's tank washer. Always use the approved tank washer provided with
the AS3000.
4. Open the bail of each yoke and mount the cylinder over the tank washer.
5. Ensure the O
2
cylinder mates to the O
2
Pin Index Safety System (PISS) connection on the
O
2
yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke
port.
6. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
7. Ensure that the N
2
O cylinder mates to the N
2
O PISS connection on the N
2
O yoke.
Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
8. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
9. Ensure that the AIR cylinder mates to the AIR PISS connection on the AIR yoke. Close the
yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
10. Open and close the cylinder valve and observe the cylinder gauge on the anesthesia
machine rise to the colored range.
2.2.10 Breathing Circuit, CO
2
Absorbent, and Liquid Vaporizer Agent
1. Attach a breathing circuit to the inspiratory and expiratory ports as detailed in the
directions for use. Attach the breathing bag and any other respiratory accessories as
described.
2. Insert the Pre-Paks one on top of the other with the wider side facing up on both Pre-
Paks. (alternatively the absorber canisters may be filled with loose fill absorbent.)
3. Install the absorber canister with a quarter turn of the lever at the bottom the absorber
assembly, this ensures a tight seal.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals