AS3000™ Service Manual 0070-10-0683 1 - 37
Theory of Operation The Breathing System
1.9 The Breathing System
1.9.1 CMV mode, inspiration
Tidal volume (T Vol.) compensates for variations in gas flow. This is to ensure that the set tidal
volume is delivered to the patient.
FIGURE 1-28 Breathing System Pneumatics, CMV Mode, Inspiration
1.9.2 CMV mode, expiration
As the patient exhales tidal volume into the expiratory limb, fresh gas enters the bellows.
Fresh gas mixes with exhaled gas after the Absorber removes CO
2
. Excess fresh gas passes
through the exhaust valve to the AGSS
FIGURE 1-29 Breathing System Pneumatics, CMV Mode, Expiration
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Datascope AS 3000 and is the answer not in the manual?

Mindray Datascope AS 3000 Specifications

General IconGeneral
BrandMindray
ModelDatascope AS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals