AS3000™ Service Manual 0070-10-0683 1 - 19
Theory of Operation Ventilator Control and Drive
1.5.1.2 Drive Gas Pressure Sensor Board
FIGURE 1-10 Drive Gas Pressure Sensor Board, Top View
FIGURE 1-11 Drive Gas Pressure Sensor Board, Bottom View
Pair Board
Pair, J1
PIN NAME FUNCTION
1. AGND Analog Ground
2. Pair Pressure of air
3. NC No connection
4. AR10V 10V
5. NC No connection