AS3000™ Service Manual 0070-10-0683 1 - 25
Theory of Operation Ventilator Control and Drive
Inspire Valve, J12
PIN NAME FUNCTION
1 - 2 24V 24V
3 - 4 Valexp inspire Valve Drive
Man/Auto Valve, J14
PIN NAME FUNCTION
1. 24V 24V
2. Valman Manual/Auto Valve Drive
Signals and Power of Sensor Board, J13
PIN NAME FUNCTION
1. A+12V Analogue +12V
2. FL_INS Import flow
3. FL_EXP Export flow
4. FM_FL Flowmeter flow
5. Pair Pressure of air
6. Paw Pressure inside air way
7. Pgas Pressure gas supply status
8. GND Ground
9. AGND Analog Ground
10. AGND Analog Ground
11. P24V Power 24V
12. P24V Power 24V
13. P24V Power 24V
14. P12V Power 12V
15. P12V Power 12
16. P5V Power 5V
17. P5V Power 5V
18. Inhale Inspire Valve
19. Man MANUAL/AUTO Valve
20. Exhale Expire Valve
21. Exhale Expire Valve
22. AGND Analog Ground
23. AGND Analog Ground
24. AGND Analog Ground
25. AGND Analog Ground