Part I Service Scenarios
Repair 90
Log Name Log Description Remarks
Peri_State Peripheral status log recording
system hardware information and
other alarm information
Running status information
about related peripherals is
recorded after the Doppler
process is started.
Monet_Log_Backend_State Backend status log
TEMPORARY_Log.log Unimportant logs containing
massive information
For example, incomplete
display of characters of a
control or thread creation
information
Monet_ERROR_Log Error log Error information in main logs
is extracted to this file,
including startup and shutdown
information.
LargeBuffDump Large file memory related log Large memory log
3.2 Troubleshooting
Note: Deal with the error based on the site situation.
3.2.1 System Power-up Failure
3.2.1.1 Involved Modules or Boards
No. Description Remarks
1. Power adapter /
2 Switch button board /
3 Main board /
4 PC module /
3.2.1.2 Information to Be Collected
No. Information to Be Collected Remarks
1. Power indicator This indicator is on the adapter. If it is on,
the 19 V output is normal. If it is off, the
output is abnormal or the indicator is faulty.
2 19 V indicator This indicator is on the main board.
3 5VSTB indicator This indicator is on the rear side of the main
board and visible after the SSD card cover
is removed.

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals