XI
Electric safety
WARNING:
Connect the power plug of this system and power plugs of the peripherals to
wall receptacles that meet the ratings indicated on the rating nameplate.
Using a multifunctional receptacle may affect the system grounding
performance, and cause the leakage current to exceed safety requirements.
Use the power cord accompanied with the system provided by Mindray.
Disconnect the AC power before you clean or
machine, otherwise, electric shock may result.
When using peripherals not powered by the auxiliary output of the
ultrasound system, or using peripherals other than permitted by Mindray,
make sure the overall leakage current of pe
ripherals and the ultrasound
system meets the requirement of the local medical device electrical regulation
(like enclosure leakage current should be no more than 500uA of IEC 60601-
1-1), and the responsibility is held by the user.
In maintenance or asse
mbly/disassembly, make sure other cables are
connected well before the battery connecting cable is connected, otherwise
the system may be damaged due to hot-plug.
Do not use this system simultaneously with equipment such as an
electrosurgical unit, high-
frequency therapy equipment, or a defibrillator,
etc.; otherwise electric shock may result.
This system is not water-proof. If any water is sprayed on or into the system,
electric shock may result.
CAUTION:
DO NOT connect or disconnect the system’s
power cord or its accessories
(e.g., a printer or a recorder) without turning OFF the power first. This may
damage the system and its accessories or cause electric shock.
Avoid electromagnetic radiation when perform performance test on the
ultrasound system.
In an electrostatic sensitive environment, don’t touch the device directly.
Please wear electrostatic protecting gloves if necessary.
You should use the ECG leads provided with the ECG module. Otherwise it
may result in electric shock.
The maximum output power of the output interface of the trolley is 240 VA.
Mechanical safety
WARNING:
When moving the system, you should first fold the LCD display, disconnect
the system from other devices (including probes) and disconnect the system
from the power supply.
Do not subject the transducers to knocks or drops. Use of a defective probe
may cause electric shock to the patient.