Product Specifications 120
1.2 Technical Specifications
1.2.1 Physical Specifications
Dimensions (D × W × H): 322±5 mm × 364±5 mm × 44±3 mm
Weight: < 4.0 kg (including the embedded battery)
1.2.2 Electrical Specifications
Voltage: 100 V to 240 V AC
Frequency: 50/60 Hz
AC input power: input current for the main unit: 2.0 A to 1.0 A; input current for the trolley: 4.4 A-
2.5 A
Battery of the main unit: 14.4 V DC
1.2.3 Environmental Specifications
Operational Conditions
˗ Ambient temperature: 0°C to 40°C
˗ Relative humidity: 20% to 85% (no condensation)
˗ Atmospheric pressure: 700 hPa to 1060 hPa
Storage and Transportation Conditions
˗ Ambient temperature: -20 to 55
˗ Relative humidity: 20% to 95% (no condensation)
˗ Atmospheric pressure: 700 hPa to 1060 hPa
1.2.4 Display Specifications
1.2.4.1 Main Display
15.6-inch display
Up/Down angle of view: ≥ 170 degrees, anti-dazzle
Left/Right angle of view: ≥ 170 degrees, anti-dazzle
Resolution: 1920 × 1080
The brightness of the main display may be adapted according to the ambient brightness.
The backlight of the display is linked in the auto adaptation state.
The contrast of the main display may be manually adjusted.
The opening-closing angle of the display is no less than 180 degrees.
1.2.4.2 Touch Screen
Size: 12.3 inches
Resolution: 1920 × 720
Hardness: 2H (2H pencil, pressure 1N/45°)

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals