Part I Service Scenarios
Installation 3
Battery capacity
6600 mAh (per battery)
1.1.2.2 Requirements of Regulator
Due to the difference of the power supply stability of different districts, please advise the user to adopt a
regulator of good quality and performance such as an on-line UPS.
1.1.2.3 Grounding Requirements
The power cord of the system is a three-wire cable. The grounding protective terminal should be
connected with grounding protective cable of the power supply system. Make sure that the protective
grounding works normally.
WARNING:
DO NOT connect this system to outlets with the same circuit breakers and fuses
that control the current of devices such as life-
support systems. If this system
malfunctions and generates an over-
current, or when there is an instantaneous
current at power ON, the circuit breakers and fuses of the building’s supply circuit
may be tripped.
1.1.3 EMI Limitation
Ultrasound devices are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transient in the air wiring. They also generate EMI. Possible EMI sources should be
identified before the unit is installed.
These sources include: medical lasers, scanners, monitors, cauterizing guns, etc. Besides, other devices
that may result in high-frequency electromagnetic interference such as mobile phone, radio transceiver
and wireless remote control toys are not allowed to be presented or used in the room. Turn off those
devices to make sure the ultrasound system can work in a normal way.
1.1.4 Packaging Specifications
Item Size Weight
Main unit package 700 × 540 × 425 mm About 25 kg (full configuration)
Traveling case 620 × 460 × 260 mm About 18 kg (full configuration)
Trolley package 630 × 630 × 1080 mm About 55 kg