Part I Service Scenarios
Maintenance 50
If you spill liquid on the main unit, disconnect the power supply, dry
the main unit, and contact your service personnel.
Never use abrasive materials (such as steel wool or silver polish),
or erosive cleaners (such as acetone or acetone-based cleaners).
Dilute and use the disinfectants according to the manufacturer's
instructions.
Check the system after disinfection. If there is any sign of damage,
remove it from use.
Follow local regulations when selecting and using the disinfectant.
Disinfectants listed in this section are used for disinfecting the
housing of the main unit and the monitor only, not for disinfecting
the probes or the trolley.
During cleaning, wear medical gloves to prevent infection.
2.1.4.1 Compatible Disinfectants
The following table lists compatible disinfectants.
Product Name Manufacturer Type
75% alcohol / Solution
99.5% isopropyl alcohol / Solution
SONO
TM
ULTRASOUND
WIPES
Advanced Ultrasound
Solutions Inc.
Wipe
Schulke mikrozid
®
Sensitive
Wipes
Schulke Wipe
Mikrobac Tissues Bode Chemie Gmbh Wipe
Sani-Cloth
®
Plus PDI Wipe
2.1.4.2 Disinfection Procedures
Note:
The system has passed puncture test and leakage current test, and thus can be
disinfected on power- on status: tap > bar in the left side of the touch screen to
enter the Exam Main Menu screen and then select to lock the system for 10
seconds to disinfect the system while the power is “On”
After cleaning, disinfect the system as follows.
1. Wear medical gloves to prevent infection.